UV Spectro Photometer |
 |
| UV Spectrophoto Meter CL-1310 |
| Wave length |
| Range |
200 to 1100 nm |
| Accuracy |
1 nm |
| Bandwidth |
2.0 nm |
| Resolution |
0.1 nm |
| Photometric |
| Operating Mode |
%T, ABS, CONC |
| Measuring Mode |
Single W/L |
| Range |
-0.3 to 2.000 ABS |
| Accuracy |
+0.005 Abs at 1.0A |
| Resolution |
0.1% T, 0.001 Abs |
| Read out |
3½ digit LED for Data and 16x1 Digit LCD for W/L |
| Grating |
1200 lines /mm |
| Source |
Tungsten Halogen Lamp 12V, 50W for 330-1100 nm, Deuterium Lamp for 200-330 nm |
| Detector |
Silicon Photo diode |
| Sample compartment |
4 Position Manual |
| Power |
230V, 50Hz |
| Accessories |
Two 10mm path lenght QUARTZ curettes |
|
| |
| UV Spectrophoto Meter CL-1320 |
| Wave length |
| Range |
190 to 1100 nm |
| Accuracy |
1 nm |
| Bandwidth |
2.0 nm |
| Resolution |
0.1 nm |
| Photometric |
| Operating Mode |
%T, ABS, CONC |
| Measuring Mode |
Single W/L Multi W/L, Scan, Time Scan |
| Range |
-0.3 to 2.000 ABS |
| Accuracy |
±0.005 Abs at 1.0A |
| Resolution |
0.1% T, 0.001 Abs |
| Read out |
20x4 alpha numeric LCD |
| Grating |
1200 lines /mm |
| Source |
Tungsten Halogen Lamp 12V, 50W for 330-1100 nm, Deuterium Lamp for 200-330 nm |
| Detector |
Silicon Photo diode |
| Sample compartment |
5 Position Auto |
| Power |
230V, 50Hz |
| Accessories |
Two 10mm path lenght QUARTZ curettes |
|
| |
|